| Madison-Omni-DP1 Specifications: |
| Relative Pressure Operating Ranges / Overload Limits / Burst Pressure: |
| Range |
Overload |
Burst |
| PSIG |
BAR |
PSIG |
BAR |
PSIG |
BAR |
0-14.5
0-29.0
0-72.5
0-87.0
0-290.1
0-725.2
0-1450.4
0-2900.8 |
0-1
0-2
0-5
0-10
0-20
0-50
0-100
0-200 |
43.5
87.0
217.6
435.1
870.2
2175.6
2900.8
4061.1 |
3
6
15
30
60
150
200
400 |
58.0
116.0
290.1
580.2
1015.3
2610.7
5076.3
8702.3 |
4
8
20
40
70
180
350
600 |
|
| Differential Pressure Range |
Customer-specified (min.: 10% of operating range; max.: operating range) |
| Accuracy |
±1% FSD |
| Operating Temperature |
32°F to 158°F (0°C to 70°C); 185°F (85°C) max. with swan neck |
| Storage Temperature |
-4°F to 176°F (-20°C to 80°C) |
| Supply Voltage |
18-30 VDC |
| Power Consumption |
< 1 W |
| Switching Values (S1 and S2) |
PNP or NPN selectable; 300 mA load in sum max.; programmable as min. or max. value; short-circuit proof; reverse-polarity proof |
| Hysteresis |
Adjustable; position of hysteresis depends on min. or max. |
| Display |
Graphical back-lit LCD display with extended temperature range (-4°F to 158°F / -20°C to 70°C); 32x16 pixels; shows value and units; LED signaling lamp with simultaneous message in display |
| Connection |
At locking plug M12x1, 5-pole |
| Protection Class |
IP67 |
| Materials: Medium Contact |
Stainless steel 1.4571, Viton, Ceramic Al2O3 |
| Materials: Electronic Housing |
Stainless steel 1.4305 (housing); tempered mineral glass (glass); cobalt samarium (magnet); POM (ring) |